Plus applications user manual  2.9.0.0be062e
Software library for tracked ultrasound image acquisition, calibration, and processing.
Probe calibration algorithm

This algorithm determines the transformation between the ultrasound image and a tracker marker rigidly attached to somewhere on the ultrasound probe.

Configuration settings

  • element.png
    vtkPlusProbeCalibrationAlgo
    • attribute.png
      ImageCoordinateFrame
      required.png
    • attribute.png
      ProbeCoordinateFrame
      required.png
    • attribute.png
      PhantomCoordinateFrame
      required.png
    • attribute.png
      ReferenceCoordinateFrame
      required.png
    • attribute.png
      OptimizationMethod. Additional optimization step to force the computed ImageToProbe matrix to be orthogonal. Usually 2D method provides slightly better results. See this paper for details: http://perk.cs.queensu.ca/contents/improving-n-wire-phantom-based-freehand-ultrasound-calibration
      optional.png
      Default = "NONE"
      • 2D Distance of actual and expected fiducial line intersection point is minimized in the image plane.
      • 3D Distance of actual fiducial point is and the fiducial line is minimized in 3D.
    • attribute.png
      IsotropicPixelSpacing Specifies if during optimization an isotropic horizontal and vertical spacing in the image is enforced. Only used if OptimizationMethod is not NONE
      optional.png
      Default = "FALSE"
  • element.png
    Segmentation: Segmentation and pattern recognition parameters. Can be checked and modified using SegmentationParameterDialogTest or fCal (FreehandClibration toolbox) applications
    • attribute.png
      ApproximateSpacingMmPerPixel
    • attribute.png
      MorphologicalOpeningCircleRadiusMm
    • attribute.png
      MorphologicalOpeningBarSizeMm
    • attribute.png
      ClipRectangleOrigin
    • attribute.png
      ClipRectangleSize
    • attribute.png
      MaxLinePairDistanceErrorPercent
    • attribute.png
      AngleToleranceDegrees
    • attribute.png
      MaxAngleDifferenceDegrees
    • attribute.png
      MinThetaDegrees
    • attribute.png
      MaxThetaDegrees
    • attribute.png
      ThresholdImagePercent
    • attribute.png
      CollinearPointsMaxDistanceFromLineMm
    • attribute.png
      UseOriginalImageIntensityForDotIntensityScore
  • element.png
    PhantomDefinition
    • element.png
      Description
      • attribute.png
        Name
      • attribute.png
        Type
      • attribute.png
        Version
      • attribute.png
        WiringVersion
      • attribute.png
        Institution
    • element.png
      Geometry
      • element.png
        Pattern
        • element.png
          Wire
          • attribute.png
            Name
          • attribute.png
            EndPointFront
          • attribute.png
            EndPointBack
      • element.png
        Landmarks
        • element.png
          Landmark
          • attribute.png
            Name
          • attribute.png
            Position

Phantom definition

We recommend using 3 N-wires as we have found this to be more robust and accurate than using just 2 N-wires but still can be robustly detected and kept in the field of view. Specification of the wire pattern is available at https://github.com/PlusToolkit/PlusDoc/raw/master/tutorials/PlusTutorialBuildingfCalPrintedPhantom.pptx. XML description can be found in the example device set configuration files.

You can add any number of N-wires with any angles between the side and middle wires (angle can be different for each N-wire) with the following conditions:

  • all N-wires shall be completely visible at the same time in the ultrasound image
  • planes of N-wires shall be parallel
  • only one N-wire is allowed on a plane (they should not be placed side-by-side, only above/below each other)

N-wires should cover about the same area of the ultrasound image (same size and location) during registration as your region of interest (e.g., if you image a 40mm diameter object 50mm from the transducer then during calibration the N-wires should cover about 40mm range, about 50mm from the transducer).

You can find the CAD models and snapshots of fCal calibration phantoms in the Plus 3D model catalog: http://perk-software.cs.queensu.ca/plus/doc/nightly/modelcatalog/

fCal-2.x is recommended for smaller field of view (up to about 10cm imaging depth). The hole positions are the same for fCal-1.x and fCal-2.x (only some text labels and marker mounting holes are added) but fCal_2.x has extra holes that allow using 3 N-wires, which greatly increases the accuracy and robustness (compared to using just 2 N-wires), therefore using fCal-1.x is not recommended.

If you image large structures then you probably need fCal 3 phantom. However, fCal-3.x is assembled from multiple pieces (to limit reduce 3D printing time and costs), and so it may be somewhat less accurate then the smaller, unibody fCal-2.x.

Example configuration file PlusDeviceSet_fCal_Ultrasonix_L14-5_Ascension3DG_2.0.xml

<PlusConfiguration version="2.3">
<DataCollection StartupDelaySec="1.0" >
<DeviceSet
Name="fCal: Ultrasonix US (L14-5 probe) + Ascension3DG tracker (Probe, Reference, Stylus) - fCal Phantom 2.0"
Description="Ascension3DG sensors should be plugged in to the Ascension3DG DriveBay mounted on Ultrasonix US in the following order from to leftmost slot (Transducer 1) to the right: 1 Probe, 2 Reference, 3 Stylus." />
<Device
Id="TrackerDevice"
Type="Ascension3DG"
LocalTimeOffsetSec="0.0"
FilterAcWideNotch="1"
ToolReferenceFrame="Tracker" >
<DataSources>
<DataSource Type="Tool" Id="Probe" PortName="0" />
<DataSource Type="Tool" Id="Reference" PortName="1" />
<DataSource Type="Tool" Id="Stylus" PortName="2" />
</DataSources>
<OutputChannels>
<OutputChannel Id="TrackerStream" >
<DataSource Id="Probe"/>
<DataSource Id="Reference"/>
<DataSource Id="Stylus"/>
</OutputChannel>
</OutputChannels>
</Device>
<Device
Id="VideoDevice"
Type="SonixVideo"
AcquisitionRate="30"
LocalTimeOffsetSec="-0.2976"
IP="127.0.0.1"
AutoClipEnabled="TRUE" >
<DataSources>
<DataSource Type="Video" Id="Video" PortName="B" PortUsImageOrientation="UF" />
</DataSources>
<OutputChannels>
<OutputChannel Id="VideoStream" VideoDataSourceId="Video"/>
</OutputChannels>
</Device>
<Device
Id="TrackedVideoDevice"
Type="VirtualMixer" >
<InputChannels>
<InputChannel Id="TrackerStream" />
<InputChannel Id="VideoStream" />
</InputChannels>
<OutputChannels>
<OutputChannel Id="TrackedVideoStream"/>
</OutputChannels>
</Device>
</DataCollection>
<CoordinateDefinitions>
<Transform From="Image" To="TransducerOriginPixel"
Matrix="1 0 0 -410
0 1 0 5
0 0 1 0
0 0 0 1"
Date="2011.12.06 17:57:00" Error="0.0" />
</CoordinateDefinitions>
<Rendering WorldCoordinateFrame="Reference">
<DisplayableObject Type="Model" ObjectCoordinateFrame="TransducerOrigin" Id="ProbeModel"
File="Probe_L14-5_38.stl"
ModelToObjectTransform="
-1 0 0 29.7
0 -1 0 1.5
0 0 1 -14
0 0 0 1" />
<DisplayableObject Type="Model" ObjectCoordinateFrame="Reference" Id="Volume"/>
<DisplayableObject Type="Model" ObjectCoordinateFrame="StylusTip" Id="StylusModel" File="Stylus_Example.stl" />
<DisplayableObject Id="PhantomModel" Type="Model" ObjectCoordinateFrame="Phantom"
Opacity="0.6"
File="fCal_2.0.stl"
ModelToObjectTransform="
1 0 0 -35.0
0 1 0 -10.0
0 0 1 -5.0
0 0 0 1" />
<DisplayableObject Type="Image" ObjectCoordinateFrame="Image" Id="LiveImage"/>
</Rendering>
<Segmentation
ApproximateSpacingMmPerPixel="0.078"
MorphologicalOpeningCircleRadiusMm="0.27"
MorphologicalOpeningBarSizeMm="2.0"
ClipRectangleOrigin="27 27"
ClipRectangleSize="766 562"
MaxLinePairDistanceErrorPercent="10"
AngleToleranceDegrees="10"
MaxAngleDifferenceDegrees="10"
MinThetaDegrees="-70"
MaxThetaDegrees="70"
MaxLineShiftMm="10.0"
ThresholdImagePercent="10"
CollinearPointsMaxDistanceFromLineMm="0.6"
UseOriginalImageIntensityForDotIntensityScore="FALSE"
NumberOfMaximumFiducialPointCandidates="20"
/> <PhantomDefinition>
<Description
Name="fCAL"
Type="Multi-N"
Version="2.0"
WiringVersion="2.0"
Institution="Queen's University PerkLab"
/>
<Geometry>
<Pattern Type="NWire">
<Wire Name="7:G1_g1" EndPointFront="30.0 0.0 20.0" EndPointBack="30.0 40.0 20.0" />
<Wire Name="8:L1_h1" EndPointFront="55.0 0.0 20.0" EndPointBack="35.0 40.0 20.0" />
<Wire Name="9:M1_m1" EndPointFront="60.0 0.0 20.0" EndPointBack="60.0 40.0 20.0" />
</Pattern>
<Pattern Type="NWire">
<Wire Name="4:G3_g3" EndPointFront="30.0 0.0 10.0" EndPointBack="30.0 40.0 10.0" />
<Wire Name="5:H3_l3" EndPointFront="35.0 0.0 10.0" EndPointBack="55.0 40.0 10.0" />
<Wire Name="6:M3_m3" EndPointFront="60.0 0.0 10.0" EndPointBack="60.0 40.0 10.0" />
</Pattern>
<Pattern Type="NWire">
<Wire Name="1:H5_h5" EndPointFront="35.0 0.0 0.0" EndPointBack="35.0 40.0 0.0" />
<Wire Name="2:L5_i5" EndPointFront="55.0 0.0 0.0" EndPointBack="40.0 40.0 0.0" />
<Wire Name="3:M5_m5" EndPointFront="60.0 0.0 0.0" EndPointBack="60.0 40.0 0.0" />
</Pattern>
<Landmarks>
<Landmark Name="#1" Position="104.3 5.0 20.0" />
<Landmark Name="#2" Position="104.3 45.0 20.0" />
<Landmark Name="#3" Position="104.3 45.0 0.0" />
<Landmark Name="#4" Position="104.3 -5.0 0.0" />
<Landmark Name="#5" Position="-34.3 45.0 15.0" />
<Landmark Name="#6" Position="-34.3 -5.0 20.0" />
<Landmark Name="#7" Position="-34.3 -5.0 0.0" />
<Landmark Name="#8" Position="-34.3 45.0 0.0" />
</Landmarks>
<Planes>
<Plane Name="Superior" BasePoint="5.5 3.0 35.0" EndPoint1="135.0 3.0 35.0" EndPoint2="135.0 55.0 35.0" />
<Plane Name="Inferior" BasePoint="5.5 55.0 0.0" EndPoint1="135.0 55.0 0.0" EndPoint2="135.0 5.5 0.0" />
<Plane Name="Outside Unmarked" BasePoint="140.0 5.0 30.0" EndPoint1="140.0 50.0 30.0" EndPoint2="140.0 50.0 5.0" />
<Plane Name="Inside Unmarked" BasePoint="125.0 40.0 30.0" EndPoint1="125.0 20.0 10.0" EndPoint2="125.0 40.0 10.0" />
<Plane Name="Outside Marked" BasePoint="0.0 40.0 30.0" EndPoint1="0.0 40.0 5.5" EndPoint2="0.0 15.5 5.5" />
<Plane Name="Outside Posterior" BasePoint="130.0 0.0 30.0" EndPoint1="20.0 0.0 30.0" EndPoint2="20.0 0.0 15.0" />
<Plane Name="Inside Posterior" BasePoint="110.0 10.0 30.0" EndPoint1="40.0 10.0 30.0" EndPoint2="30.0 10.0 10.0" />
<Plane Name="Outside Anterior" BasePoint="130.0 60.0 30.0" EndPoint1="20.0 60.0 30.0" EndPoint2="20.0 60.0 5.0" />
</Planes>
<References>
<Reference Name="#1" Position="104.3 5.0 20.0" />
<Reference Name="#2" Position="104.3 45.0 20.0" />
<Reference Name="#3" Position="104.3 45.0 0.0" />
<Reference Name="#4" Position="104.3 -5.0 0.0" />
<Reference Name="#5" Position="-34.3 45.0 15.0" />
<Reference Name="#6" Position="-34.3 -5.0 20.0" />
<Reference Name="#7" Position="-34.3 -5.0 0.0" />
<Reference Name="#8" Position="-34.3 45.0 0.0" />
</References>
</Geometry>
</PhantomDefinition>
<VolumeReconstruction OutputSpacing="0.5 0.5 0.5"
ClipRectangleOrigin="0 0" ClipRectangleSize="820 616"
Interpolation="LINEAR" Optimization="FULL" CompoundingMode="MEAN" FillHoles="OFF" />
<fCal
PhantomModelId="PhantomModel"
ReconstructedVolumeId="Volume"
TransducerModelId="ProbeModel"
StylusModelId="StylusModel"
ImageDisplayableObjectId="LiveImage"
NumberOfCalibrationImagesToAcquire="200"
NumberOfValidationImagesToAcquire="100"
NumberOfStylusCalibrationPointsToAcquire="200"
RecordingIntervalMs="100"
MaxTimeSpentWithProcessingMs="70"
ImageCoordinateFrame="Image"
ProbeCoordinateFrame="Probe"
ReferenceCoordinateFrame="Reference"
TransducerOriginCoordinateFrame="TransducerOrigin"
TransducerOriginPixelCoordinateFrame="TransducerOriginPixel"
TemporalCalibrationDurationSec="10"
FixedChannelId="VideoStream"
FixedSourceId="Video"
MovingChannelId="TrackerStream"
MovingSourceId="ProbeToTracker"
DefaultSelectedChannelId="TrackedVideoStream"
FreeHandStartupDelaySec="5" />
<vtkPlusPivotCalibrationAlgo
ObjectMarkerCoordinateFrame="Stylus"
ReferenceCoordinateFrame="Reference"
ObjectPivotPointCoordinateFrame="StylusTip" />
<vtkPlusPhantomLandmarkRegistrationAlgo
PhantomCoordinateFrame="Phantom"
ReferenceCoordinateFrame="Reference"
StylusTipCoordinateFrame="StylusTip"
DetectionTimeSec="1.0"
StylusTipMaximumDisplacementThresholdMm="1.8" />
<vtkPlusProbeCalibrationAlgo
ImageCoordinateFrame="Image"
ProbeCoordinateFrame="Probe"
PhantomCoordinateFrame="Phantom"
ReferenceCoordinateFrame="Reference" />
</PlusConfiguration>