Plus applications user manual  2.9.0.0be062e
Software library for tracked ultrasound image acquisition, calibration, and processing.
Data collection diagnostics (DiagDataCollection)

Acquire raw, non-synchronized, non-interpolated tracking and video data and compute some basic metrics for diagnostics. In addition to the output on the screen two files are generated in the output directory containing the frame numbers and filtered and unfiltered timestamps (in [date]_[time].VideoBufferTimestamps.txt and [date]_[time>]TrackerBufferTimestamps.txt).

Examples

DiagDataCollection.exe --config-file=..\..\PlusLib\data\ConfigFiles\Test_PlusConfiguration_VideoNone_FakeTracker_PivotCalibration_fCal.xml --input-acq-time-length=10

Command-line parameters reference

Help: 
  --acq-channel-ids opt opt ..  Identifiers of the output channels that are
                                recorded. If not specified then all channels
                                are recorded.

  --acq-time-length=opt  Length of acquisition time in seconds (Default:
                                60s)

  --config-file=opt  Name of the input configuration file.

  --help  Print this help.

  --output-seq-file-prefix=opt  Filename prefix for the recorded output
                                channels (Default: Diag)

  --verbose=opt  Verbose level (1=error only, 2=warning, 3=info,
                                4=debug, 5=trace)